MENU LIMITS

Menu- Level 1
Level 2

7.7.1. Limits for internal control

Menu functions
The transparent thermostat CT 72 / P can only be set to maximum 60.00 °C.
The configuration does not permit higher values.
Press the
key if parameter is to be retained. Correction function for
parameters or values (prior to ENTER).
Parameter level
0 ... 100 %
0 ... 100 %
SETPOINT MAX / MIN – Maximum and minimum setpoint
Restriction of the adjustable temperature range
The limitation of the operating temperature range effects the temperature
setting in the menu with the key
Only setting of working temperatures which lie within the determined limits is
possible
Existing settings for SETPNT 1, -2, -3, as well as those for >OVERTMP< and >
SUBTMP < (refer to page 70), are automatically deferred into the limit range.
Setting range: -94,90 °C ... +200,0 °C
SET MAX > SET MIN
Interchange of values is not possible.
Set maximum heating / cooling
The heating and cooling capacity of the unit are adjustable. 100 % corresponds
to the technical specification of the equipment.
Setting range:
HEAT MAX –
COOLING MAX –
Integer digits flash, set by pressing
+
Decimal digits flash, set by pressing
+
Integer digits flash, set by pressing
+
Decimal digits flash, set by pressing
+
The value flashes, set by pressing
+
The value flashes, set by pressing mit
+
.
0 to 100 % in 1 % steps
0 to 100 % in 1 % steps