Error message with ticker:
> INTERNAL HARDWARE ERROR-CALL SERVICE <
Other errors
Error in A/D converter
Error message with ticker:
> EXCESS TEMPERATURE PROTECTOR ALARM-CHECK ADJUSTMENT <
Excess temperature protector defect.
The protection temperature is below the set working temperature setpoint.
Set the protection temperature to a higher value.
Error message with ticker:
> SAFETY SENSOR ALARM-CALL SERVICE <
The cable of the excess temperature protector has been disconnected or short-
circuited
By quickly switching off and restarting the unit the alarm is cancelled.
60
If the error occurs once more after the restart, a remote diagnosis is required.
45
15
4 s
30
Error message with ticker:
> CONFIGURATION ERROR-PRESS OK<
The configuration of the circulator does not correspond with its current application.
Press the
In this case please call the Technical Service or an authorized dealer.
Disturbances that are not indicated.
The electronic pump motor is overload-protected by an electronic current limiter. If
viscosity of the bath fluid is or becomes too high, the motor stops running.
Mains circuit breakers (resettable) 15 A
15
15
A
A
M
M
P
P
Troubleshooting guide / error messages
key for a non-recurring, automatic change of the configuration.
91