2 . General information about the bracket for the via GO
The bracket for the via GO is designed for combination with standard and lightweight wheelchairs with a frame tube diameter of 25
mm / 1 inch. Depending on the seat width, there are different pre-mounted brackets.
For wheelchairs to which this standard bracket cannot be attached, there are optional extensions depending on the wheelchair. For
these cases, please observe the mounting instructions enclosed with these extensions.
3 . Preparing the mounting of the bracket for the via GO
3.1 Preparing the wheelchair and getting parts ready
• Align the wheelchair.
• Tighten the two knee lever brakes.
• Get the following parts ready:
- Bracket for the drive unit
- Meter stick/tape measure
- Allen wrench SW5
- Torque wrench
3.2 Determine the height of the bracket
• Measure the distance A from the floor to the lower frame tube [5] of the wheel-
chair.
Distance A
(from the floor the lower frame
tube of the wheelchair)
170 to 210 mm
210 to 220 mm
220 to 230 mm
230 to 240 mm
A
240 to 250 mm
250 to 260 mm
Example:
The distance A from the floor to the lower edge of the lower frame tube [5] of the
wheelchair is 232 mm. Therefore, screw the adapter [3] to the outer clamp [2] at
position 4 and 5.
1
2
3
4
5
6
7
Position
Tightening
of the clamping
torque of the
screws
1 and 2
8 Nm
2 and 3
8 Nm
3 and 4
8 Nm
4 and 5
8 Nm
5 and 6
8 Nm
6 and 7
8 Nm