CHECKING THE LEVEL ACCURACY OF THE VERTICAL BEAM
0,5 m
0,5 m
1. Hang an approximately 2 m plumb line on
a wall.
2. After the plumb line has settled, mark
point a1 on the wall, behind the plumb
B
B
line, near the plumb cone.
180°
180°
b1, b2
b1, b2
3. Set up the cross laser on a tripod or on a
flat surface in front of the wall at a
distance of approximately 2.5 m.
4. Switch ON the self-leveling mode, and
press the mode button to project the
vertical beam towards the plumb line.
5. Turn the laser so that the vertical beam
will merge with the plumb line below the
hanging point.
6. Mark point a2 on the wall, in the middle of
the vertical beam at the same height as
a1.
7. The distance between a1 and a2, should
be no more than 0.75 mm
ENGLISH
4
This procedure requires a room of with
approx. 5m distance between floor and
ceiling.
1. Set up the cross laser on a tripod.
2. Switch ON the self-leveling mode
3. Mark the upper plumb dot on the
4. Mark the lower plumb dot on the
a1
a1
5. Rotate the laser level 180°, position
6. Mark the center of the upper plumb
7. The distance between points a1
a2
a1
a2
a1
8. The distance between a1 and a2,
a1
a1
CHECKING PLUMB ACCURACY
and press the mode button to
switch to the plump function.
ceiling as point a1 (see the
illustration).
floor as point b.
it so that the center of the plumb
dot is on the point b, which has
already been marked, and allow it
to level.
dot on the ceiling as point a2.
and a2 marked on the ceiling is an
indication of the actual deviation of
the laser level from the ideal plumb
line.
If the measuring distance is 5m,
the maximum permissible deviation
is:
5 m x (+/- 0.3 mm/m) x 2 = +/ - 3 mm.
should not be more than 6 mm.
a1
a1
5 m
5 m
b
b
a1
a2
a1
a2
180°
180°
b
b
12