7. Interface
DATA-BITS: Defined value (8)
STOP-BITS: Defined value (1)
PARITY-BITS: Defined value (N)
7.4

Commands for serial interface

All interface commands are preceded by a colon and completed with the carriage return
(CR). Command and parameters must be separated by a space. By adding a question
mark to the appropriate action the parameter can be read rather than changed. Received
commands are acknowledged with SELECT, null or false commands with ERROR.
Command
Auto mode
:saaz <0 | 1>
:acy <1 ... 100>
:asd <1 ... 100>
:asu <1 ... 100>
:ate <0 ... 10000>
:ath <1 ... 10000>
:atp <1 ... 10000>
:atr <1 ... 10000>
:ats <1 ... 10000>
Interface output
:o <0 | 1>
WIKA operating instructions, model CPC2000
Answer function
Auto zero before each cycle (phase 4)
0 --> off
1 --> on
Cycles to be carried out
1 ... 100 --> number of cycles
Steps Down
1 ... 100 --> number of steps to reach the end point
Steps Up
1 ... 100 --> number of steps to reach the end point
Delay time at the end point (phase 13)
1 ... 10000 --> time in seconds
Hold time (phase 10 and 16)
1 ... 10000 --> time in seconds
Pause time (phase 18)
1 ... 10000 --> time in seconds
Tolerance band
1 ... 10000 --> tolerance in 0.01 % FS (full scale) of the instrument
measuring range. Once the instrument has been controlled within this
tolerance band for 1 second, the hold time expires.
Start delay (phase 2)
1 ... 10000 --> time in seconds
Output status information over interface
0 --> off
1 --> on
EN
27