6. Commissioning, operation
SETTINGS submenu:
Go to the SETTINGS menu item with the
In the submenu, go to the zero point adjustment item and confirm with SELECT.
Under the point, AUTO-ZERO, the automatic zero point adjustment for the MEAS
EN
mode can be activated or deactivated.
Operation
On leaving the menu, the zero point of the internal sensor is adjusted, dependent upon
the setting. The instrument then begins to measure the subsequent pressures.
Using the ZERO button in MEAS mode, the instrument's zero point can be compensated
at any time.
CAUTION!
If the connected pressure should exceed the permissible highest loading
(125 %), the internal sensor will be closed off using a magnetic valve. The
display will then show ERROR. Using the SELECT button, the instrument
can be closed off and opened up freely.
6.3

Operating mode 'AUTO'

In AUTO mode, a pressure profile can be stored. This is especially useful when multiple
pressure sensors or pressure switches are to be calibrated with the same values. The
mode also enables a defined profile to be traversed several times in succession.
20
/
button, confirm with SELECT.
Full scale value
Pressure unit
Measured value
Set point in % of the full scale
value
Battery status
Operating mode
WIKA operating instructions, model CPC2000