WIKA CPC2000 Manual De Instrucciones página 29

Controlador portátil de bajas presiones,
Ocultar thumbs Ver también para CPC2000:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

7. Interface
Command
Menu settings
:smm <a | c | m >
:ssw <1 ... 100>
:swm <a | c | m | z l v
| s>
:szc <0,1>
:szi <1 ... 60>
Additional commands
:sce <0,1>
:pi? <CR>
:pj? <CR>
:pk? <CR>
WIKA operating instructions, model CPC2000
Answer function
Define the operating modes of the instrument
a --> start automatic mode
c --> start control mode
m --> start measure mode
Step size
1 ... 100 --> step size from operating the up and down buttons in %
Define the operating modes of the instrument
a --> start automatic mode
c --> start control mode
m --> start measure mode
(only available in CONTROL and MEASURE modes)
z --> ZERO
(only available in CONTROL mode)
l --> LEAK test
l --> back to CTRL mode (like LEAK button)
v --> VENT (vent the entire system)
(only available in AUTOMATIC mode)
Zeroing before entry into control mode
0 --> off
1 --> on
Zeroing interval
1 ... 60 --> time in minutes
Set Communication Interface Echo
0 --> Echo Off
1 --> Echo On
On: Default. Sent command is returned, following the response, the
command will be concluded with "OK".
Off: Only the pure response will be transmitted.
Reading current ACTUAL pressure with unit
Example:
-0.05;mbar
Reading current ACTUAL pressure
Example:
-0.05
Reading current unit as text
Example:
Mbar
EN
29