Installation; Electromagnetic Environment - Gima DIATERMO 108 Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

GIMA SpA
Inspect the unit for damages during transport. The claims for possible damages will be
accepted only in case they are immediately communicated to the carrier; the damages that are
found must be written down and presented to LED SpA or to your own retailer. If the unit is
returned to the LED SpA or to your own retailer, it is necessary to use the original equipment's
package or another equivalent one, to guarantee the safety during the transport.
Unpack the equipment and carefully study the documentation and operating instruction
supplied. Mains voltage, indicated above the inlet, must agree with the local mains voltage
(mains voltage frequency: 50-60 Hz). The correct voltage setting is selected by turning the
voltage selector when available. Insert the correct fuses in the module referring to the value
written on the label. (see 4.6.2).
Connect mains cable to a mains outlet having good hearth connection
OPERATION OF THE EQUIPMENT WITHOUT EARTH CONNECTION IS FORBIDDEN
The unit must be installed on a level surface, with dimension, at least, correspondent to those of
the base of the unit itself. Around the unit must be left a space of 25cm, at least.
The equipment must be installed and be used in the electromagnetic environment in paragraph
3.1.
Connect the mains cable to the mains socket on the rear panel of the unit.
Connect the foot switch to the connector on the front panel.
Connect handle on the black buckle.
Connect the cable to the patient plate.
Let unit work in dry environment only. Any verified condensate must be let evaporate before
putting in operation the unit. Don't exceed the temperature environment or the allowed
moisture.
When the unit is switched on, through the on/off switch on the rear panel, after having checked
the internal parameters, it will work with the function and the power level utilized during the
last switching (when the unit is switched for the first time the level will be 00).
Before using the unit, it is necessary connect the cable to the patient plate. Single plate
electrodes and split plate electrodes can be. If the value of the impedance is acceptable, the OC
indicator light will stop flashing.
Pre-set the current for the cut (CUT or BLEND) or the coagulation (COAG or MICRO), press
the foot switch to deliver the pre-set current.

3.1 ELECTROMAGNETIC ENVIRONMENT

Warning:
The equipment must be installed and used in the environment described as below, it must be
installed with cables it is equipped, which are verified the electromagnetic compatibility.
The use of accessories other than those specified, may result in increased emission or decreased
immunity level.
Manuale d'Istruzioni / Instruction's Manual / Manual de Instrucciones
DIATERMO 108

3. INSTALLATION

Temperature: from 10°C to 40°C
Relative moisture: from 30% to 75%
Pressure: from 70 kPa to 106 kPa
MA254IGBEa
32