GIMA SpA
The mains input module incorporates the mains input socket, the main input voltage selector and in-
line mains fuses.
WARNING: before switch on the unit, operator has to verify that requested mains voltage
corresponds to the voltage available from the electrical net.
4.6.2 V
S
OLTAGE
ELECTOR
Before powering the unit, it is necessary to preset the correct mains voltage by properly set the
voltage selector. Proceed as follows:
A-B) Extract the fuse holder drawer from the power module.
(C) Insert the fuses making reference to the following chart:
Mains Voltage
Mains Voltage
(D) Extract and rotate the detachable part in way to read the correct voltage in the (E) window –
reinsert the fuse holder in the module.
Connect mains cable to a mains outlet having good hearth connection.
The unit is factory preset to 230Vac.
Manuale d'Istruzioni / Instruction's Manual / Manual de Instrucciones
DIATERMO 108
115 Vac Delayed fuse T4A - 5 x 20 mm
230 Vac Delayed fuse T2A - 5 x 20 mm
MA254IGBEa