4.4 PRESELECTION OF THE DELIVERABLE CURRENT

The deliverable current for the surgical operations can have preselected through push button for:
4.4.1 C
C
UT
URRENT
The best current for the cut is the pure sinusoidal wave without modulation that means with duty-
cycle 100%. Such current, proper for cut without coagulation.
4.4.2 C
OAGULATED
The coagulated-cut current (BLEND) it is suited for coagulated cut when a deep coagulation
together the cut is desired. This current is made by sine current e suited for the cut associated to low
voltage current suited for coagulation. With this blending, a current suited for cut coagulated in
absence of eschar and carbonization is obtained
4.4.3 C
OAGULATION
The modulated current COAG it is characterized by good property of surface coagulation behaving
at the time it probable production of eschar and partial carbonization of the tissue. The advantage of
this type of coagulation resides in the rapidity with which the effect is gotten.
4.4.4 M
-C
ICRO
OAGULATION
Modulated current MICRO is suitable for little blood-vessel coagulation. The coagulation process is
more checked and exact in this case.

4.5 CONNECTORS

4.5.1 F
-
OOT
SWITCH
On the left part of the front panel it is pedal connector. The foot-switch is allowed to control the
output current delivering of the operative condition.