Maintenance; General; Cleaning Of The Cabinet; Cleaning And Sterilisation Of The Accessory Items - Gima DIATERMO 108 Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

GIMA SpA

6.1 GENERAL

No user adjustable parts are within the equipment, either for calibration or service purposes. The
equipment housing must not be opened: the warranty is invalidated by any unauthorized entry into
the unit. In the event any repair or adjustment work being necessary, the whole equipment should be
returned to the LED S.p.A. Service Centre 04010 APRILIA (LT) - ITALY, or to a other Authorised
Centre, together with a description of the fault.
Maintenance work by the user is mainly the cleaning of the exterior of the cabinet, cleaning and
sterilisation of the accessory items and checking of the equipment before each use. Carrying out
function and safety check for verification of the parameters is demanded to specialised technical
people.

6.2 CLEANING OF THE CABINET

Switch the equipment off completely and disconnect the mains supply before any cleaning is
undertaken. Clean the outside of the cabinet with a damp cloth. No chemical should be used; a mild
non abrasive cleanser may be used when necessary.

6.3 CLEANING AND STERILISATION OF THE ACCESSORY ITEMS

The best thing to do is to use only one time use accessories and discard them after use. Since some
of the accessory items are to be used more than once it is mandatory to clean carefully and sterilise
those accessories before the new use. The best way to clean and sterilise the reusable items is to
follow the direction of the supplier of each item.

6.4 GUIDE TO THE SOLUTION OF THE PROBLEMS

In case of problems before all it is advised to check for the correct installation of the unit and for the
correct connection of the accessories.
Problems
The equipment
doesn't switch on.
Alarm OC always
active.
The unit doesn't
respond to the
command of
activation.
Error Code 91
Error Code 92
Error Code 93
Error Code 94
Manuale d'Istruzioni / Instruction's Manual / Manual de Instrucciones
DIATERMO 108

6. MAINTENANCE

Probable Cause
Interruption or absence of the main
feeding
Interruption or lack of contact on the
neutral electrode circuit
Breakdown of the handpiece or of
the pedal
Wrong connection of the handpiece
or of the pedal
Current delivery control activated
during switching on
Error in the management board
Error in the management board
Error in the data conversion circuit
Solution
Verify the connection of the main cable.
Verify the fuses and replace them, where
necessary, with new ones of the propre type.
Check the connection of the cable to the neutral
electrode.
Replace the cable of connection of the neutral
electrode.
Replace the handpiece or the pedal.
Verify the connection of the handpiece or of the
pedal.
Disconnect the pedal and switch on the unit
again.
Call for Service
Call for Service
Call for Service
MA254IGBEa
43