Problems
Error Code 95
Error Code 99
6.5 REPAIRS
High frequency cables and electrode holder handle cannot be repaired. Always substitute a damaged
part with a new one.
6.5.1 F
S
USE
UBSTITUTION
Before substituting the fuse, disconnect the unit from the mains system.
Only use fuse of the kind 5x20; they must have those characteristics: T2A (slow) (230Vac mains
voltage), T4A (115Vac mains voltage), proceed as follows:
(A-B) Extract the fuse holder drawer from the power module.
(C) Insert the fuses making reference to the following chart:
Mains Voltage
Mains Voltage
(D) Extract and rotate the detachable part in way to read the correct voltage in the (E) window –
reinsert the fuse holder in the module.
6.6 CHECKING OF THE EQUIPMENT BEFORE EACH USE
Each time the use of the electrosurgical equipment is planned a check of the most important safety
aspects has to be implemented considering at least the following:
•
Check the integrity of cords, connections, wires breakage, etc.
•
Assure that all the electrical equipment is properly grounded.
•
Assure that all the accessories that should be used are available and sterilised.
•
Check, by disconnecting the reference electrode cable, the functioning of the OC light and
sound alarm warning.
§
Check, by activating the CUT and COAG power switch, the functioning of the emission
sounds warnings.
6.7 FUNCTION AND SAFETY CHECK AND TEST
At least once a year, the biomedical engineering department or other qualified personnel should do
the following check and test:
•
Check of the connectors and mains supply cord conditions.
•
Visual check of the mechanical protections.
Manuale d'Istruzioni / Instruction's Manual / Manual de Instrucciones
DIATERMO 108
Probable Cause
Error of the reference voltage value
Error in the output power activation
circuit
115 Vac Delayed fuse T4A - 5 x 20 mm
230 Vac Delayed fuse T2A - 5 x 20 mm
Solution
Verify the main voltage
Call for Service
Call for Service
MA254IGBEa