For the models equipped with FIEM MINILED THERMOSTAT
1. Product overview
A
B
2. Key functions
3. Functionality description
MINILED is able to control with ON-OFF method or PID algorithm the temperature (see paragraph 7) and with mode
ON-OFF the incubation humidity.
At power on, a welcome message will appear briefly followed by firmware version of the board.
MINILED will prompt the current temperature and by pressing the arrow UP or DOWN the relative humidity.
By pressing the SET button you can modify the temperature and humidity value already set.
In case temperature or humidity probe fail, or in case one of them isn't connected, in the relative visual mode will prompt
ERR and the output will not controlled.
MINILED has in addition two different menus, one for the user, where you can customize the basic functions, and one
technical that permit to setup the machine and adjust the probe.
During start-up, menu for setting the type of temperature control.
4. User setup
After you turn-on MINILED by pressing the
close to temperature icon start blinking and display show the
reference temperature.
Now by pressing
or
• T. incubazione [5 .. 75] (default 30.0)
This parameter define the temperature that the machine must be maintened during the incubation period. The range of
values you can set is between 5 and 75 °C or equivalent in F.
You can save the value by pressing the
Now the GREEN LED close to humidity icon start blinking and display show the humidity percentage to be observed For
modify the humidity parameter pressing
• Umidità incubazione [10 .. 99] (default 40)
This parameter allows you to set the relative humidity to be maintained during the incubation period. The range of values
that can be set in this field is between 10% and 99%.
After editing the values MINILED automatically exits the edit menu if no buttons are pressed for 15 seconds by saving the
value shown.
5. Technical setup
By simultaneously pressing
pressing
, button, the modify of the parameter occur by pressing
To confirm the modify of the value press again
MINILED will exit automatically after 15 seconds if no keys are pressed by saving the value shown.
List of parameters showed by pressing
• Unit [C,F] (default °C) - Setting the temperature units Celsius or Fahrenheit
• t.CAL [-3,0 .. +3,0 °C] (default 0,0 – step di 0,1 °C)
t.CAL [-5,4 .. +5,4 F] (default 0,0 – step di 0,1 F) - It allows to calibrate the temperature read by the probe, in Celsius or
Fahrenheit depending on the setting assigned at the previus point.
• H.Cal [-5 .. +5] (default 0 – step di 1%) - It allows to calibrate the humidity value read by the probe.
C
Shows the inside temperature
Setting temperature and humidity reference values
Shows the inside humidity
+
Technical menu for the calibration parameters, measurement units
+
During start-up, menu for setting the type of temperature control
button, RED LED
you can change the set temperature.
button, display will show the humidity value.
or
button.
and
you enter the technical menu. Its possible to read all parameters in sequence by
.
button
F
D
A. Power switch
B. Light switch
C. 3 digits display with dot separation value
G
D. temperature LED
E. humidity LED
F. selection button <arrow up> (temperature display)
E
G. setting button
H
H. selection button <arrow down> (humidity display)
and
.