retailer in the moment of equivalent new device purchasing, one against one. Disposing of the product separately prevents possible negative
consequences for the environment and for health, deriving from inadequate disposal. It also allows the recovery of materials of witch it's made up in
order to obtain an important saving of energy and resources and to avoid negative effects to the ambient and health. In case of abusive disposal of
device by user, will be applied administrative endorsements in compliance with current standard.
This equipment should only be utilised on an horizontal working surface.
WARNING: Ensure that the IN marker on the filter is on the side facing the collection jar lid and fitted into the "VACUUM".
A wrong connection causes immediate destruction in case of contact with sucked liquids.
Filter assembling
Suction Pump
Inlet
•
Connect the short silicon tube 1, with antibacterial filter 2, to the suction connector 3. The other tube, with one end connected to the filter must
be connected with the other end to the jar's connector 4 where has been fixed the red float (security float).
•
When the 90% of the volume of the jar is reached there is the activation of the security float (the float close the aspiration connector on the jar)
to avoid liquid penetration inside the device. The device must be used on a plan of horizontal operation.
•
Connect the long silicon tube 5 to the other jar's lid connector 6
•
Connect the other end of the long silicon tube 5 to the probe plastic connector 7
•
Connect the power cord to the device than connect the plug to the electrical mains supply.
•
Unscrew the jar's lid and fill the jar 1/3 full or ordinary water (this for an easy cleaning operations and an rapid reaching of the functionality
vacuum) then rescrew the lid on the jar correctly.
•
Push switch 8 on position I to start suction
•
Once finished push switch on O position and unplug
•
To extract the accessories and start with cleaning it.
The dual use version is available on request.
Continuous use can be enabled by activating the illuminated switch.
Alternatively, the pedal control can be used for aspirating on command, while leaving the operator's hands free. If you're using the pedal version,
insert the plug of the pedal power cord into the appropriate socket located next to the equipment, push the second switch on the position II, so start
and stop the suction operation by pressing relasind the pedal.
Using FLOVAC® disposable collection system: Before connecting the disposable collection system, remove the white ring fitted on the tank
holder for a more comfortable insertion of the same container.
INSTRUCTIONS
FLOW DIRECTION
IN / Fluid Inside
IN
16
Jar Air Tube