Descargar Imprimir esta página

Dräger Oxylog VE300 Instrucciones De Uso página 17

Ocultar thumbs Ver también para Oxylog VE300:

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 29
4.3
Mounting the ceiling holder
1. Make sure that the surface is sturdy enough.
2. Drill the 4 fixation holes according to the dimension drawing.
3. Screw the ceiling holder to the ceiling with 4 M6 screws.
4. Screw the mounting plate to the ceiling holder with the 4 screws.
5. Screw the wall holder to the mounting plate with the 4 screws.
6. To change the position of the ceiling holder, rotate the ceiling holder horizontally.
5
Inserting Oxylog VE300
5.1
Inserting Oxylog VE300
Risk of patient injury
Bent or pinched hoses and lines may compromise sufficient supply to the patient.
► When inserting the device in the device holder, make sure that the device ports are
1. Place the groove on the side of the device on the middle of the lower holding bar.
2. Slightly press the holding bracket up and secure the upper part of the device behind the
3. Make sure that the device is properly secured.
6
Removing Oxylog VE300
6.1
Removing Oxylog VE300
Risk of patient injury
Accidentally removed or torn hoses and lines may compromise sufficient supply to the
patient.
► Before removing the device, make sure that all hoses and lines can move freely.
1. Slightly press the holding bracket up and remove the upper part of the device from the
2. Remove the device from the lower holding bar. While doing so, take care with any freely
|
Instructions for use
Device holder for Oxylog VE300
Oxylog VE300 can only be inserted without a carrying system.
Oxylog VE300 can be inserted on both sides of the device holder.
WARNING
easily accessible and that no hoses or lines are bent or pinched.
holding bar. The device can also be clicked into place by pressing it in firmly without
lifting the holding bracket.
WARNING
holding bar.
moving hoses and lines.
Inserting Oxylog VE300
|
English
17

Publicidad

loading