Set up and connect the unit
8.1
Operating modes
Operating mode
Switched off
Switched on
Safety switch-off with
wake-up function
Ultrasonic cleaning
Ultrasonic cleaning
complete
9
Set up and connect the unit
9.1
Notes on installation site
The unit is designed to be set up in commercially used premises such as laboratories, medical practices,
etc.
• The installation site must be well ventilated.
• The permissible ambient conditions [} 9] must be met.
• The unit must be installed on a sturdy, horizontal, non-slip, moisture-resistant work surface with the
following characteristics:
– The surface must be non-flammable.
– There must be sufficient clearance above the unit to allow the cover and objects being cleaned to
be removed safely.
– There must be an area available next to the unit for setting down the basket and inserts.
• Electrical connection conditions:
– Splash-proof earthed socket near the unit
(cable length approx. 1.5 m).
– The power connection must be protected by a residual current circuit breaker (RCCB).
– Ensure that the power supply [} 9] required to operate the unit is available.
• Wastewater connection requirements:
– Select an installation site located close to a sink that can be used to empty the unit.
CAUTION! Use heat-resistant material.
14 / 24
Description
The cleaning duration rotary knob is turned to OFF. All LEDs are off.
The cleaning duration rotary knob is turned to ON, or a cleaning duration of
1–30 minutes is set. The LED of the set ultrasound mode lights up.
The unit is switched on. Automatic safety switch-off occurs after 8 hours of
inactivity or in case of a brief interruption of the power supply. The LEDs turn
off. The unit can be switched on again by pressing a button or turning the
rotary knob.
Ultrasonic cleaning was started by pressing the Start/Stop button. The LED
for cleaning duration flashes during ultrasonic cleaning.
You can press Start/Stop again to terminate ultrasonic cleaning at any time.
The LED for cleaning duration is off.
A brief audible signal is emitted at the end of the cleaning duration. The LED
for cleaning duration lights up.