Guidance and manufacturer´s declaration – electromagnetic emission – for all EQUIPMENT AND SYSTEMS
The TheraFace Mask is intended for use in the electromagnetic environment specified below. The customer or the user of TheraFace Mask should assure that it is used in such an environment.
EMISSIONS TEST
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage function / flicker emissions
IEC 61000-3-3
The TheraFace Mask is intended for use in the electromagnetic environment specified below. The customer or the user of the TheraFace Mask should assure that it is used in such an environment.
IMMUNITY TEST
Electrostatic discharge (ESD)
IEC 61000-4-2
Electrostatic transient / burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short interruptions and voltage variations on power
supply input lines
IEC 61000-4-11
Power frequency (50/60 Hz magnetic field
IEC 61000-4-8
NOTE
U
is the a.c. mains voltage prior to application of the test level.
T
GUIDANCE AND MANUFACTURER´S DECLARATION – ELECTROMAGNETIC EMISSION
COMPLIANCE
Group 1
The TheraFace Mask uses RF energy only for its internal function. There for, its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
Class B
The TheraFace Mask suitable for use in all establishments, including domestic establishments and those directly connected to the public low-
voltage power supply network that supplies buildings used for domestic purposes.
Class A
Complies
GUIDANCE AND MANUFACTURER´S DECLARATION – ELECTROMAGNETIC IMMUNITY
IEC 60601 TEST LEVEL
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air
±2kV for power supply lines
±1kV for input/output lines
±1kV differential mode
±2kV common mode
0% UT; 0,5 cycle g) At 0°, 45°, 90°, 135°, 180°, 225°, 270°and 315°
0 % UT; 1 cycle and 70 % UT; 25/30 cycles Single phase: at 0°
0 % UT; 250/300 cycle
30 A/m
ELECTROMAGNETIC ENVIRONMENT - GUIDANCE
COMPLIANCE LEVEL
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air
±2kV for power supply lines
±1kV differential mode
45°, 90°, 135°, 180°, 225°, 270° and 315°
0 % UT; 1 cycle and 70 % UT; 25/30 cycles Single phase: at 0°
0 % UT; 250/300 cycle
30 A/m
ELECTROMAGNETIC ENVIRONMENT - GUIDANCE
Floors should be wood, concrete or ceramic tile. If floors are
covered with synthetic material, the relative humidity should be
at least 30 %.
Mains power quality should be that of a typical commercial or
home environment.
Mains power quality should be that of a typical commercial or
home environment.
Mains power quality should be that of a typical commercial or home
environment. If the user of the TheraFace Mask requires continued
operation during power mains interruptions, it is recommended that
the TheraFace Mask be powered from an uninterruptible power
supply or a battery.
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commercial or home
environment.