Sterilisation Table; Filling Of The Water Tank - Tecno-gaz ANDROMEDA PLUS xp Manual De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

PAR. 7
CYCLES
C1
134°C
C2
134°C
C3
121°C
C4
121°C
C5
Special
134°C
The data of this table are indicative; the choice of the sterilisation cycle must be based on the data
supplied by the producer of the object you want to sterilise.
PAR. 8
TURN ON THE DEVICE PRESSING THE MAIN SWITCH.
Connect the suitable supplied hose to the loading plug and insert the other end (the one with the filter)
inside the demineralised or distilled water tank.
NOTE: The use of feedwater with contaminants at levels exceeding those given in the table above can
greatly shorten the working life of the device, with damage at the components of the autoclave and
then invalidate the manufacturer's warranty.
Pressing the button 14A, the pump is automatically turned on. By sucking, the pump fills the tank.
The pump supplies water for 120 seconds; after reaching the maximum level, it is turned off.
In case during these 120 seconds the pump does not reach the maximum level, the operator must press the
button 14A up to switching off the pump in automatic mode (for the maximum reached level).
The autoclave does not work in case the tank level in on MIN. as signalled by the LED.
The attempt to start the autoclave with MIN. level state is visualised with the message "ER2" on the display
"TIME/MESSAGE.
STERILIZATION TABLE
TIME OF
TIME OF
STERILIZATION
DRYING
4 minutes
12 minutes
9 minutes
12 minutes
18 minutes
12 minutes
19 minutes
12 minutes
4 minutes
2 minutes
FILLING OF WATER TANK
MAXIMUM LOAD FOR
EACH TRAY (Kg)
1
1
1
1
1
22
WORKING PRESSURE
(Relative Bar)
2.1÷2.2
2.1÷2.2
1.1÷1.2
1.1÷1.2
2.1÷2.2
DPTA208 Rev_2