POTENTIAL FOR ELECTROMAGNETIC
INTERFERENCE
To avoid inaccurate results caused by electromagnetic interference
between electrical and electronic equipment, do not use the device near
a cell phone or microwave oven. For most wireless
communication devices, it is recommended to
maintain a distance of 10.8 feet (3.3m) in order to
avoid electromagnetic interference.
This device complies with Part 15 of the FCC Rules.
Operation is subject to the following two conditions:
(1) this device may not cause harmful interference,
and (2) this device must accept any interference
received, including interference that may cause
undesired operation.
FEDERAL COMMUNIATIONS COMMISSION COMPLIANCE STATEMENT
Changes or modifications to this equipment not expressly approved by the
manufacturer could void the user's authority to operate the equipment.
This equipment has been tested and found to comply with the limits for
a Class B digital device, pursuant to Part 15 of the FCC Rules. These
limits are designed to provide reasonable protection against harmful
interference in a residential installation. This equipment generates, uses
and can radiate radio frequency energy and, if not installed and used in
accordance with the instructions, may cause harmful interference to radio
communications. However, there is no guarantee that interference will
not occur in a particular installation. If this equipment does cause harmful
interference to radio or television reception, which can be determined by
turning the equipment off and on, the user is encouraged to try to correct
the interference by one or more of the following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the equipment and receiver.
• Connect the equipment into an outlet on a circuit different from
that to which the receiver is connected.
• Consult the dealer or an experienced radio/TV technician for help.
37
ELECTROMAGNETIC COMPATIBILITY (EMC)
• Guidance and manufacturer's declaration – electromagnetic emissions
The device is intended for use in the electromagnetic environments listed below, and should only be used in such
environments:
Emissions test
Compliance
Electromagnetic environment – guidance
RF emissions CISPR 11
Group 1
RF energy is used only to maintain device's operation.
Therefore, its RF emissions are so low that it's not likely to
RF emissions CISPR 11
Class B
cause any interference in nearby electronic equipment.
Not
Harmonic emissions IEC
The device is suitable for use in all establishments, including
61000-3-2
Applicable
domestic establishments, and those directly connected to
the public low-voltage power supply network that supplies
Voltage fluctuations/ flicker
Not
buildings used for domestic purposes.
emissions IEC 61000-3-3
Applicable
• Guidance and manufacturer's declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environments listed below, and should only be used in such
environments:
Immunity test
IEC 60601 test level
Compliance level
Electrostatic
± 6 kV contact
± 8 kV contact
discharge (ESD)
± 8 kV air
± 15 kV air
IEC 61000-4-2
Power frequency
3 A/m
30 A/m
(50/60 Hz)
magnetic field
IEC 61000-4-8
• Recommended separation distances between portable and mobile RF communication equipment and the device.
The device is intended for use in an electromagnetic environment where radiated RF disturbances are under control. User
can help prevent electromagnetic interference by keeping the device at a minimum distance from portable and mobile RF
communications equipment (transmitters). Below table details the maximum output power of transmitter:
Rated maximum
Separation distance according to frequency of transmitter m
output power of
150 kHz to 80 MHz
80 MHz to 800 MHz
transmitter W
Not Applicable
d = 1.2
0.01
NA
0.1
NA
1
NA
10
NA
100
NA
38
Electromagnetic environment–
guidance
The relative humidity should be at
least 5 %.
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
Recommended separation distance
I
r = --------- (m)
188
where I is the current in amperes in a
power bus or an appliance wire and r is
the recommended separation distance
between your device and the power bus
or appliance wire, in meters (m).
800 MHz to 2.5 GHz
d = 2.3
0.12
0.23
0.38
0.73
1.2
2.3
3.8
7.3
12
23