E4 - Data transmission
Bluetooth function is turned
error symbol shown on the
OFF, on your mobile device.
display
Bluetooth function is turned
OFF, on the blood pressure
monitor.
Mobile device does not pair
with the blood pressure
monitor.
The App on the mobile
device is not open.
The blood pressure monitor
and mobile device are out of
transmitting range.
Unexpected loss of electri-
cal/mechanical integrity.
EP error symbol shown on
System error.
display
Turn Bluetooth ON, on your
mobile device, and try again.
Turn the Bluetooth function
ON, on your blood pressure
monitor (see page 19), then
try again.
Make sure Bluetooth is ON,
on the mobile device and the
blood pressure monitor, and
try again.
Requires compatible mobile
device. This blood pressure
monitor is compatible with
mobile devices that are run-
ning iOS 5.0 or newer, or
AndroidTM 4.3 and newer.
For a complete list, go to
www.walgreens.com/con-
nectapp.
Make sure the app is open
on the mobile device, and
try again.
Make sure the mobile device
and blood pressure monitor
are within the acceptable
range of 32 feet (10 meters).
Make sure your last reading
is stored in memory and the
App is open and try again.
See page 32.
Remove batteries, re-insert,
and try again.
Return the device to your
local distributor or importer.
Measure again.
41
SPECIFICATIONS
Power Source:
(2)1.5V DC "AAA" Alkaline batteries
Measurement
Oscillometric
Method:
Measurement
Pressure: 40~280 millimeters Mercury (mmHg)
Range:
Pulse: 40~199 beats/minute
Pressure: ±3 mmHg
Accuracy:
Pulse: ±5% of reading
Pressure Sensor:
Semi-conductor
Inflation:
Pump driven
Deflation:
Automatic Pressure release valve
Memory Capacity: 120 memories for each User (240 total)
Auto-shut-off:
1 minute after last button operation
Operation
Temperature: 41°F~104°F (5°C~40°C)
Humidity: 15% ~ 93% RH
Environment:
Storage/
Temperature: -13°F~158°F (-25°C~70°C)
Transportation
Humidity: Less than 93% RH
Environment:
42