Descargar Imprimir esta página

Stryker CUB FL19H Manual De Uso página 19

Cuna pediátrica
Ocultar thumbs Ver también para CUB FL19H:

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

A
Display
A
lb/kg
B
C
Zero
Change equipment
D
Weigh/On
E
Memory
F
W W e e i i g g h h i i n n g g a a p p a a t t i i e e n n t t
C C A A U U T T I I O O N N
• Always remove external equipment from the product litter when you zero the scale or weigh a patient.
• Always zero the scale at least once a week to maintain scale accuracy.
To weigh a patient:
1. Press and release the W W e e i i g g h h / / O O n n button (E) to turn the scale on ( Scale control panel option (page 14)).
N N o o t t e e - - Three dashes will flash on the display for 10 seconds followed by a weight reading.
2. Press and hold the Z Z e e r r o o button (C) for two seconds ( Scale control panel option (page 14)).
N N o o t t e e - - Patient must be moved from product to zero the scale.
3. Release the Z Z e e r r o o button and wait for a zero weight to display.
4. Place the patient on the product and press and release the W W e e i i g g h h / / O O n n button. Three dashes will flash on the display for
10 seconds followed by the patient weight.
N N o o t t e e - - The
icon may display when using the scale and indicates the product is unstable and unable to weigh.
Stabilize the product and repeat the process.
1900-009-005 Rev AB.0
B
D E
Displays patient weight or error messages
Changes the unit of measurement
Sets the scale to zero and erases the previous memory
Adds or removes equipment without affecting patient weight
Activates the scale
Stores the most recent patient weight or displays the stored weight
C
F
15
EN

Publicidad

loading