8. Turn the knob (A) clockwise to tighten the knob.
9. Reverse steps to disconnect the IV caddy from the product.
To store the IV caddy, place the IV caddy in the storage tray or secure the caddy in the storage clip.
P P o o s s i i t t i i o o n n i i n n g g t t h h e e t t w w o o - - s s t t a a g g e e p p e e r r m m a a n n e e n n t t l l y y a a t t t t a a c c h h e e d d I I V V p p o o l l e e o o p p t t i i o o n n
W W A A R R N N I I N N G G
• Always make sure that the IV pole is positioned so that the patient cannot reach or manipulate the IV pole.
• Do not use the IV pole as a push/pull device. Product damage may occur.
C C A A U U T T I I O O N N
• Do not use a foot end fixed IV pole on a product with the scale option.
• Do not load the IV pole above the safe working load of 40 lb (18 kg).
• Do not load an IV pole hook above the safe working load of 11 lb (5 kg).
You can purchase the product with the two-stage IV pole option permanently attached at the head end, foot end, or both
ends of the product. The IV pole is equipped with a telescopic pole that extends to provide a second height position. You
can fold and store the IV pole when not in use.
To position the two-stage IV pole (Figure 13):
1. Lift and pivot the pole from the storage position.
2. Push the IV pole down until the IV pole locks in place.
3. To raise the height of the IV pole, pull up on the telescoping portion (A) until the pole locks in place at the fully raised
position.
4. Rotate the IV hangers (B) to the desired position and hang the IV bags.
5. To lower the IV pole, hold the telescoping portion of the IV pole, turn the latch (C), and lower the telescoping portion.
N N o o t t e e - - Always make sure that the IV pole is at a low height to pass safely through door openings when you transport a
patient.