F F i i g g u u r r e e 1 1 4 4 – – P P o o s s i i t t i i o o n n i i n n g g t t h h e e t t h h r r e e e e - - s s t t a a g g e e p p e e r r m m a a n n e e n n t t l l y y a a t t t t a a c c h h e e d d I I V V p p o o l l e e
A A t t t t a a c c h h i i n n g g a a n n d d p p o o s s i i t t i i o o n n i i n n g g t t h h e e r r e e m m o o v v a a b b l l e e I I V V p p o o l l e e ( ( F F A A 6 6 4 4 1 1 3 3 5 5 ) )
For use with the premium accessory bracket, not compatible with the standard accessory bracket.
W W A A R R N N I I N N G G - - Do not use the IV pole as a push/pull device. Product damage may occur.
C C A A U U T T I I O O N N
• Always remove external equipment from the product litter when you zero the scale or weigh a patient.
• Do not load the IV pole above the safe working load of 40 lb (18 kg).
To attach and position the removable IV pole (Figure 15):
1. Insert the IV pole into a socket at the head end or foot end of the product.
2. Turn the knob (A) counterclockwise and pull up on the telescoping portion (B) until you reach the desired height.
3. Turn the knob (A) clockwise to lock the telescoping portion in place.
N N o o t t e e - - Always make sure that the IV pole is at a low height to pass through door openings when you transport a patient.