1-3 Electrical Specifications
Accuracy is given as ±(% of reading + counts of
least significant digit) at 23°C ± 3°C, with
relative humidity Less than 80% R.H.
ACA specifications are ac coupled, true R.M.S.
The crest factor may be up to 3.0
as 4000 counts.
For non-sinusoidal waveforms, Additional
Accuracy by Crest Factor (C.F.):
Add 3.0% for C.F. 1.0 ~ 2.0.
Add 5.0% for C.F. 2.0 ~ 2.5.
Add 7.0% for C.F. 2.5 ~ 3.0.
Position Error of Clamp :
Add ±2% of LCD reading.
Max ±4% near the jaw opening.
Adjacent Conduction Influence of Jaw : <0.08A/A
AC Current
Range Resolution
60.00A
300.0A
All specifications are valid from 10% to 100%
of each range.
LCD displays 0 counts when the reading < 10
counts.
Low-pass Filter
Range
60.00A
300.0A
All specifications are valid from 10% to 100%
of each range.
LCD displays 0 counts when the reading < 10
counts.
Cut-off Frequency (-3dB) : Approx. 160Hz
Attenuation Characteristic : Approx. -24dB / Oct
0.01A
0.1A
Resolution
0.01A
0.1A
Accuracy
(50 ~ 60 Hz)
±(2.0% + 5D)
(50Hz~60Hz)
±(3.5%+5D)
±(3.5%+5D)
7
Accuracy
EN