Communications Interface
14 Communications Interface
14.1 Instructions for Setup
RS232 serial communications are available. They are accessible via the DB-9
connectors on the rear of the chiller. Refer Inadmissible Operating Conditions to
avoid any safety hazards.
A Terminal Emulator or other comparable device will need to be connected to
the DB-9 to allow command to be entered.
Terminal Settings:
none
14.2 Commands and responses
Command
OUT_MODE_01˽1
OUT_MODE_05˽0
IN_MODE_05
OUT_SP_00
˽XXX,XX
IN_SP_00
IN_PV_00
IN_PV_01
STATUS
VERSION
106
Baud Rate 115200, Data – 8-bit, Parity – none, Stop – 1 bit, Flow Control -
Description
Start tempering
Stop tempering
Read whether tempering is
active or not
Change the target
temperature to XXX.XX
Read the target
temperature
Read the current
temperature
Read the current
heating/cooling power
Read the current status
Read the unit name and
firmware version
General Response
none
none
0: tempering stopped
1: tempering started
none
XXX.XX
XXX.XX
-100 (100% cooling) ... +100 (100%
heating)
See table below
JULABO TE400 VERSION 1.01