8.2
Disconnect Hoses
The coolant hoses are connected to the unit at the coolant inlet and coolant
outlet connections, labeled with respective symbols.
Requirements
Unit prepared for maintenance. Refer to Preparing the Unit for Maintenance.
Coolant cooled to the ambient temperature.
Required Tools and Materials
Absorbent cloth
Bonding agent
Protection caps
Procedure
1. The coolant inlet and outlet fittings have quick disconnects installed.
The coolant hoses are now disconnected from the unit.
8.3
Connecting Power
Requirements
The unit construction provides protection against the risk of electrical shock
Power Cord with C13 connector
Use cable rated for 10A 250V with IEC320-C13 receptacle. The customer side
Procedure
1. Cable with a 'C13' Connector should be connected to the IEC power
2. Turn the Power Switch ON.
3. When the Power Switch is ON, the Power Switch should light up as well as
Disconnect the hoses from coolant inlet and outlet fittings on the back of the
unit.
by grounding appropriate metal parts. The protection will not function unless
the power cord is connected to a properly grounded outlet. It is the user's
responsibility to assure a proper ground connection is provided.
of the cable must follow required standard for the country of installation
connection on unit as shown below.
the LCD Panel.
Installation Procedure
85