The Theragun RELIEF is intended for use in the electromagnetic environment specified below. The cus-
tomer or the user of the Theragun RELIEF should assure that it is used in such an environment.
Immunity test
Voltage dips, short interrup-
tions and voltage variations
on power supply input lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
12
Guidance and manufacturer´s declaration – electromagnetic immunity
IEC 60601 test level
0 % UT; 0,5 cycle g) At 0°, 45°, 90°,
135°, 180°, 225°, 270° and 315°
0 % UT; 1 cycle and 70 % UT; 25/30
cycles Single phase: at 0°
0 % UT; 250/300 cycle
30 A/m
NOTE
UT is the a. c. mains voltage prior to application of the test level.
Compliance
Electromagnetic environ-
level
Mains power quality should be
that of a typical commercial or
home environment. If the user of
the Theragun RELIEF requires
continued operation during
N/A
power mains interruptions, it is
recommended that the Theragun
RELIEF be powered from an
uninterruptible power supply or
a battery.
Power frequency magnetic fields
should be at levels characteristic
30 A/m
of a typical location in a typical
commercial or home environment.
ment -guidance