Descargar Imprimir esta página

Therabody THERAGUN RELIEF Manual página 14

Ocultar thumbs Ver también para THERAGUN RELIEF:

Publicidad

Guidance and manufacturer´s declaration – electromagnetic immunity – for EQUIPMENT and SYSTEM
EN
The Theragun RELIEF is intended for use in the electromagnetic environment specified below. The cus-
tomer or the user of the Theragun RELIEF should assure that it is used in such an environment.
Immunity
test
Radiated RF
10 V/m
IEC 61000-4-3
80 MHz to 2.7 GHz
385MHz-5785MHz
Test specifications
for ENCLOSURE
PORT IMMUNITY
to RF wireless
communication
equipment (Refer
to table 9 of IEC
60601-1-2:2014)
14
Guidance and manufacturer´s declaration – electromagnetic immunity
IEC 60601 test
Compliance
level
10 V/m
80 MHz to 2.7 GHz
385MHz-
5785MHz Test
specifications for
ENCLOSURE
PORT IMMUNITY
to RF wireless
communication
equipment (Refer
to table 9 of IEC
60601-1-2:2014)
Electromagnetic environment -guidance
level
where p is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in metres (m).b
Field strengths from fixed RF transmitters, as determined
by an electromagnetic site survey,a should be less than the
compliance level in each frequency range.
Interference may occur in the vicinity of
equipment marked with the following symbol:

Publicidad

loading