9. Product Specifications

9.1 Delivery form

The IVOMAT IP3 consists of the following components:
– Apparatus with air filter control
– Non-return valve
– 1.5 m compressed air hose
– 1.2 m water drain hose with canister cap
– Canister hose with sound absorber; polymerization
container in pressure chamber, and 3 object clips
secured to removable lid
– 1 return water container
– 1 replenishing cup

9.2 Technical data

Power supply: Single-phase alternating current
Standard version:
– 220 V / 50 Hz
– 240 V / 50 Hz
Acceptable voltage fluctuations: +10 % to –15 %
Power consumption: 1030 W
Electrical fuses:
100–118 V: 12.5 A, slow. Diameter: 6.3 x 32 mm
220 V: 6.3 A, slow. Diameter: 5 x 20 mm
240 V: 5 A, slow. Diameter 5 x 20 mm
Compressed air:
Operating pressure (already set): approx. 6 bar = 6.10 Pa
(Tolerance: green area of the scale)
System pressure: 6 to max. 12 bar = 6 to max. 12.10 Pa
Hose with an inside diameter of 6 mm
Dimensions of the closed apparatus:
Width: 312 mm, Depth: 302 mm
Height: 220 mm
Useful dimensions of pressure chamber:
Diameter: 112 mm, height: 98 mm
Max. water capacity 0.78 dm
Weight: 9.75 kg
Safety information:
EN 61010, Part 1, EMC tested
3
(Liter)

9.3 Acceptable operating conditions

Acceptable temperature range
+5 °C to +40 °C (+41 °F to +104 °F)
Acceptable altitude:
The apparatus has been tested for use at altitudes of up
to 2000 m
Atmospheric pressure of 500 mbar to 1060 mbar
9.4 Acceptable transportation and storage
conditions
Acceptable temperature range
–20 °C to +55 °C (-4 °F to +131 °F)
Maximum relative humidity: 80 %