ROUTING CABLES THROUGH THE EXOSKELETAL
EN
SHOULDER JOINT (FIGURE 4)
In the exoskeletal version, both input and output wires are routed through a channel between the yoke
plate and the humeral plate.
▪ (Figure 4A) To electrodes, switches, etc.
▪ (Figure 4B) Lock output wire
▪ (Figure 4C) To input board
1.
A hole entering the socket must be provided for the Axis output wire.
2.
Input wires travel from the input control (electrodes, switches, etc) to the input connector
board. The Axis output wire travels from the shoulder joint to the input connector board.
3.
Attach cables to input board.
Info: For more information on electrode fabrication, see Espire Elbow
Fabrication Instructions.
ROUTING CABLES THROUGH THE ENDOSKELETAL
SHOULDER JOINT (FIGURE 5)
▪ (Figure 5A) To electrodes, switches, etc.
▪ (Figure 5B) Lock output wire
▪ (Figure 5B) To input board
In the endoskeletal version, both input and output wires are routed through the center of the shoulder joint.
1.
A hole entering the socket must be provided for the Axis output wire.
2.
Run all cables from socket, through center of shoulder joint, down outside of pylon, to elbow
input board.
3.
Fasten cables to pylon with tape.
4.
Insert cables through one of two openings on the endo adapter.
5.
Attach cables to input board.
Info: For more information on endo adapter assembly, see Espire Elbow Endo
Adapter Kit Instruction Sheet
Info: Use of a cosmetic cover is recommended to encase and protect cable
wires.