1.0 Introduction
The BUFFALO FILTER PlumeSafe TURBO Auto-Sense Smoke Evacuation System is designed to
remove smoke and aerosols and noxious odors produced during, but not limited to, laser and
electrosurgical procedures.
The PlumeSafe TURBO Auto-Sense Smoke Evacuation System has been designed with a high
suction, high flow rate centrifugal action pump. The ultra quiet motor is used to draw the smoke
plume from the surgical site through the vacuum tubing/hose into the ViroSafe filter where the
surgical smoke is processed by 4 stages of filtration. Built into a single housing, this disposable filter
is used to simplify the installation and removal during filter changes. The filter is completely enclosed
to protect the healthcare personnel from potential contamination during filter changes and is totally
filter life automated via RFID technology.
2.0 Inspection
The BUFFALO FILTER PlumeSafe TURBO-AS Smoke Evacuation System has been thoroughly
tested and inspected before shipment from the factory. Please check the unit before setup to insure
that no damage has occurred in transit. If damage is evident, please contact BUFFALO FILTER
customer service at 1-800-343-2324 (USA) or 1-716-835-7000 (international).
3.0 Operational Information
The BUFFALO FILTER PlumeSafe TURBO-AS Smoke Evacuation System complies with UL/IEC
60601.1 electrical standard in the following systems:
100v- 132v 50/60 Hz
220v- 240v 50/60 Hz
4.0 Warnings and Cautions
Please note that all Cautions and Warnings should be read and understood before any use of this
equipment. This smoke evacuation system should only be operated by personnel who have been
properly trained in its operation.
Warnings:
The Buffalo Filter ViroSafe Filters are completely disposable. Please handle these
disposable filters according to your local codes, regulations, and institution policy.
This device is not intended for evacuation of fluid. If fluid is expected to be aspirated to
the ViroSafe Filter, fluid collection devices must be installed with the vacuum hose
assembly. Failure to install a fluid collection device could cause filter blockage and
electrical damage.
DO NOT OPERATE THIS DEVICE IN THE PRESENCE OF EXPLOSIVE GASES.
EQUIPMENT NOT SUITABLE FOR USE IN THE PRESENCE OF FLAMMABLE
ANESTHETIC MIXTURE WITH AIR OR WITH OXYGEN OR NITROUS OXIDE.
REFER ROUTINE SERVICING TO QUALIFIED BIOMEDICAL TECHNICAL PERSONNEL.
There are no user serviceable components in the PlumeSafe TURBO Auto-Sense
Smoke Evacuation System.
Use only with the power cord provided and always plug into a grounded outlet.
®
®
1