*ESE
Function:
Set or return the bits in the Event Status Enable Register (ESER). The ESER
enables the Standard Event Status Register (SESR) to be summarized on bit 5
(ESB) of the Status Byte Register (SBR).
Syntax:
*ESE <NR1>
*ESE?
<NR1>
is in the range from 0 through 255.
Returns:
<NR1>
is a number from 0 to 255 that indicates the decimal value of the binary
bits of the ESER.
Examples:
*ESE 65
sets the ESER to binary
If the ESER contains the binary value
value of 130.
*ESR? (query only)
Function:
Return and clear the contents of the Standard Event Status Register (SESR).
Syntax:
*ESR?
Returns:
<NR1>
is a number from 0 to 255 that indicates the decimal value of the binary
bits of the ESER.
Examples:
If the ESER contains the binary value
value of 198.
*IDN? (query only)
Function:
Return the unique identification code of the EST.
Syntax:
*IDN?
Returns:
<string>
includes Manufacturer, model #, serial # and firmware version.
Hipot Electrical Safety Testers Models H111 and H216
0100 0001
.
1000 0010
1100 0110
*ESE?
, the
will return the
*ESR?
, the
will return the
21