5.3
Operation Method
5.3.1 Main Display
Test/Step
Mode
1
:
0
1 V A C
I m a x = 0 1
Trip Current Limit
5.3.2 Table of Parameters
Test / Step
Mode
Output
Voltage / Current
ARC
Status
42
Output Voltage/Current
V = 5
.
0 0 m A
(Actual settings may differ at power-up)
Up to 10 tests can be stored; each test may include up to
16 steps.
The first number indicates the test number while the
second number indicates the step. (e.g. 3:1 means that the
1st Step in test number 3 is active)
VAC: AC Withstanding voltage test
VDC: DC Withstanding voltage test
IR:
Insulation Resistance test
CNT: Continuity check
The available functions will differ depending on model.
AC: Output voltage (0.100 to 5.000kV)
DC: Output voltage (0.100 to 6.000kV)
IR:
Output voltage (500V or 1000V)
CNT: Output current (100mA)
If the ARC function is enabled, the sign "]" will appear
when an ARC occurs during the test.
MENU: Browse and check the steps of each test
EDIT:
Edit parameters
SAVE:
Save parameters
UTIL:
Browse and check system utility
READY: Ready for test
TEST:
Testing
PASS:
The result of the test is good
FAIL:
The result of the test is bad
STOP:
Stop the test
Hipot Electrical Safety Testers Models H111 and H216
.
0 0 0 k V
T E S T :
Ramp/Test Time
ARC
Status
* R E A D Y
0 0 0
.
0
s