5.4.2 Edit/Save the Steps
1.
Select a test and step to be edited (see § 5.4.1).
2.
Press the EDIT/SAVE button to enter the EDIT mode.
���������
3.
Use the rotary knob to select the mode parameter.
Choose from VAC, VDC, IR or CNT.
4.
Use the FIELD button to move to the output/voltage selection. Use the arrow
buttons (on either side of the Field button) to select the digit you wish to
change. Change the underlined digit using the rotary knob.
5.
Repeat step 4 to adjust the value to the exact number (e.g. 3.250kV).
6.
After setting all parameters, press the EDIT/SAVE button to enter the "SAVE"
mode. After the step is saved, the tester will automatically return to the EDIT
mode.
Hipot Electrical Safety Testers Models H111 and H216
�����
� � � �
� � �
� � � � � � � � �
45