• Press the FIELD button again to enter the next field which is "test time". Use
the rotary knob to adjust the desired testing time. The arrow buttons adjust the
knob's resolution (0 to 999.9s).
• Press FIELD button again to return the first field which is "test mode" again.
• Press EDIT/SAVE button to save all the parameters.
5.8.2 Continuous AC/DC Withstanding Voltage Test (C
• The C
/C
AC
DC
ditional Hipot tester, you can use the rotary knob and arrow keys to adjust the
output voltage during testing (10V step max.).
• All the parameters of C
time. The testing of C
ramp is still adjustable).
5.8.3 Insulation Resistance Test (IR)
• Press the MENU button to enter the "MENU" mode, then use the rotary knob
and arrow buttons to select a test.
• Press the EDIT/SAVE button to enter the "EDIT" mode. The cursor enters the
"test mode" field. Use the rotary knob to select the "IR" mode.
• Press the FIELD button to edit the next field which is "output voltage". Use the
rotary knob to adjust the desired output voltage to 500V or 1000V.
Hipot Electrical Safety Testers Models H111 and H216
test is available only on the step "COM" (see §5.7). Like the tra-
/C
AC
DC
/C
AC
DC
Functions of the FIELD Button
Output voltage
Measurement upper limit
Rmax=XXXXMΩ
Measurement lower limit
Rmin=XXXXMΩ
TEST=X.XXXs
are the same as V
is continuous (even if you program a time, the
Test mode
IR
ê
V=X.XXXkV
ê
ê
ê
Test time
AC
/V
, except the testing
AC
DC
/C
)
DC
53