3. Ex marking
Ensure that the CPG1500 can be used for its application in accordance with the media
compatibility, the process specifications and the approvals.
Approval
Marking
ATEX
Exi
IECEx
Exi
CSA
Exi
Only to 6,000 bar
[85,000 psi]
AEx
Approval ratings ATEX and IECEx
Gases and mist
→ Type examination certificates can be found online at www.wika.com.
WIKA additional operating instructions, model CPG1500 (Ex i)
Zone 0 gas
Zone 1 mounting to zone 0 gas
Zone 1 gas
Zone 0 gas
Zone 1 mounting to zone 0 gas
Zone 1 gas
Zone 2 gas
Class I, zone 0
Class I, division 1
Class I, zone 0
Class I, division 1
Installation in zone 1 (EPL Gb)
II 1G Ex ia IIC T4 Ga
II 1/2G Ex ia IIC T4 Ga/Gb
II 2G Ex ia IIC T4 Gb
T4 at -10 ... +50 °C
Ex ia IIC T4 Ga
Ex ia IIC T4 Ga/Gb
Ex ia IIC T4 Gb
Ex ia IIC T4 Gc
T4 at -10 ... +50 °C
Ex ia IIC T4 Ga
Groups A, B, C and D T4
T4 at -10 ... +50 °C
AEx ia IIC T4 Ga
Groups A, B, C and D T4
T4 at -10 ... +50 °C