TECHNICAL SPECIFICATIONS OF DEVICE
Model:
NEBULAIR combined RF7-2
Operating pressure (with neb.):
Delivery:
(1)
MMAD:
(2)
Respirable fraction < 5 μm (FPF)
(2)
Aerosol Output (AO)
(2)
Aerosol Output Rate (AOR)
(2)
Residual volume
(1)
(1) Data collected according to Flaem's internal procedure.
(2) In vitro characterization performed by TÜV Rheinland Italia S.r.l. in collaboration with the University of Parma
according to EN ISO 27427. Further details are available upon request.
APPLIED PARTS
Type BF applied parts are:
Weight:
ENVIRONMENTAL CONDITIONS
Operating conditions:
Ambient temperature
Relative air humidity
Atmospheric pressure
Storage and transport conditions:
Ambient temperature
Relative air humidity
Atmospheric pressure
DURATION
Model: NEBULAIR
(Compressor unit)
Model: RF7-2
(Nebuliser and accessories)
1.30 bar.
0.65 ml/min approx.
:
0.120 mg/min
patient accessories (B3, B4, B5, B6)
2.400 kg
Between +10°C and +40°C
Between 10% and 95%
Between 69 KPa and 106 KPa
Between -25°C and +70°C
Between 10% and 95%
Between 69 KPa and 106 KPa
Service life 2000 hours.
The expected average life is 1 year, however it is advisable to replace the
nebuliser every 6 months during intensive use (or sooner if the nebu-
liser is clogged) to ensure maximum therapeutic effectiveness.
9
Speed selector B2.4
in pos Max
in pos Min with valve
0.29 ml/min approx.
2.81 μm
75.7%
0.418 mg
1.08 ml
2.61 μm
78.2%
0.431 mg
0.093 mg/min
1.14 ml