Final inspection
• The lift unit and the trolley must be installed and first-time inspected by authorized personnel (inspector) in accordance
with the installation instructions provided by Handicare AB. Periodic inspection of the equipment should be undertaken
at least once a year. Original spare parts should be used and service/maintenance of the equipment should be in
accordance with the user manual.
• Check to ensure that no parts have been left in the packaging.
• Check to ensure that the lift has no signs of damage from transportation.
• Check emergency stop function by pressing the emergency stop, and then pressing either the up or the down button.
If nothing happens when the up or down buttons are pressed, the emergency stop is functioning properly.
• Check the entire length of the strap for signs of damage and inspect seams for wear. Press the up button and run the
strap all the way up. Then, press the down button and run the strap all the way down.
• Test lift function by lifting a person (not a user ) using an approved lifting sling. At the same time, check both the user
manual and electrical emergency lowering function with someone in the lift. See section on Emergency lowering.
• Check to ensure that the rail system is equipped with end stops.
• Check the hand control cable for signs of wear.
• Check to ensure that hand control markings coincide with lift functions.
• The user/attendant shall always pay attention to and report any unusual noises or vibration of the equipment.
NOTE!
Before the lift is used for the first time it must be charged for at least 3 hours. See section on charging the lift unit.
RisePorto is delivered with the emergency stop activated to ensure that the battery does not drain during transportation
and storage. Activate the lift unit by pulling the emergency stop outwards until a click sound is heard.
Keep the user manual where it is accessible to users of the product.
Always keep the key for mechanical emergency lowering close by the lift unit. Everybody who uses the lift unit shall be
informed where the key is kept.