Using the product
Important information
• The RisePorto lift unit can be moved between different rail systems by the attendant. The lift unit must only be used in
combination with the Trolley 360 RisePorto.
• The attendant shall be trained professional with thorough understanding of the equipment and lifting of disabled
persons. It is important that proper accessories and techniques are used by the attendant and it is also very important
that a user never is left unattended in a lifting situation.
• It is strongly prohibited to modify the original RisePorto product.
• Warranty does not apply if repairs or alterations are made to the lift by personnel who are not authorized.
• RisePorto must not come in direct contact with water.
• The lift must not be charged in a wet room.
• To ensure optimal function, the lift must be inspected regularly. See section on Maintenance.
• Under no circumstances may safe working load be exceeded. See section on technical information and the product
label on the lift.
• Lifting accessories must be properly trial fitted and tested in relation to the user's needs and functional ability.
• articular care must be taken when using the equipment near strong sources of interference, such as diathermy
apparatus, so that e.g., diathermy cables and similar equipment are not placed near the device. When in doubt, consult
your equipment technician or the supplier.
Before use
Ensure that the product is correctly mounted in the trolley.
Check rail end stops.
Check safety latch function.
Check lifting movement.
Inspect the lift for signs of wear and damage.
Check the hand control cable for signs of wear.
Check to ensure that hand control markings coincide with lift functions.
Check battery charge level.
SystemRoMedic
8
M A N U A L
TM