Immunity to interference levels by near fields of wireless HF communication devices
Radio service
GSM 1800
CDMA 1900
GSM 1900
DECT
LTE bands 1, 3, 4, 25
UMTS
Bluetooth
WLAN 802.11 b/g/n
RFID 2450
LTE band 7
WLAN 802.11 a/n
Electromagnetic compatibility (EMC)
Interference immunity measurements on supply input
Immunity to interference by rapid transient bursts – AC
voltage grid
IEC 61000-4-4:2012
± 2 kV
100 kHz repetition frequency
Immunity to interference, line-line
IEC 61000-4-5:2005
± 0.5 kV, ± 1 kV
Immunity to interference by surges, line-earth
IEC 61000-4-5:2005
± 0.5 kV, ± 1 kV ± 2 kV
Immunity to interference, line-conducted disturbances
induced by high-frequency fields – AC voltage grid
IEC 61000-4-6:2013
3 V
0.15 - 80 MHz
6 V
ISM frequency bands
0.15 - 80 MHz
80 % AM at 1 kHz
Immunity to interference due to voltage dips, short inter-
ruptions and voltage fluctuations
IEC 61000-4-11:2004
0% U
for 0.5 cycle
T
0% U
for 1 cycle
T
70% U
for 25/30 cycles (50/60 Hz)
T
0% U
for 250/300 cycles (50/60 Hz)
T
2210200496L41 1911V006
Product description
Frequency band
MHz
1700 - 1990
2400 - 2570
5100 - 5800
Conforms
Conforms
Conforms
Conforms
Conforms
EN-
US
Test level
V/m
28
28
9
15