Descargar Imprimir esta página

OHAUS Ranger 2000 Serie Manual De Instrucciones página 11

Publicidad

Idiomas disponibles

Idiomas disponibles

Ranger™ 2000 Series
5.
The scale will display [CL.lImM] (clear Check limits).
6.
Press No to use the stored check limits and proceed to step 9.
Press Yes to establish new check limits. The scale will then display [Set. Lo].
7.
Press Yes to view the "Low" limit value. Press Yes to accept or No to edit
the "Low" limit value. The stored value then displays with the first digit
highlighted [000.000 %]. Repeatedly press No until the desired number
appears. Press Yes to accept and highlight the next digit.
Repeat until all the digits are correct. Press Yes to accept the "low" limit
value, [Set. Hi] will be displayed.
Repeat the same procedure to accept or edit the "high" value.
8.
9.
If required, place an empty container on the pan and press Tare.
10. Place sample material on the pan or in the container. If the sample weight is
under the target percentage range, the yellow LED will light. If the sample is
within the target percentage range, the green LED will light. If the sample is
over the target percentage range, the red LED will light.
Note: Press Function to view the low and high check limits.
For numeric settings, please see section 4.2 for details.
3.6.3 Check Count
Set Checkweigh to Count in the Mode menu. Use this mode to compare the
quantity of items to a target quantity range.
1.
Press and hold Mode until [CHeCk] (Check) is displayed. [CLr.PwW] (clear
APW) will then display.
If no APW exists, the scale will display [Put.10], proceed to step 5.
2. Press No to use the stored APW and proceed to step 5.
3. Press Yes to establish a new APW. Scale will now display [Put.10]. Press No
or Back to toggle the choices (5, 10, 20, 50 or 100).
4.
Put the indicated number of pieces on the pan then press Yes to calculate
the APW.
5. The scale will display [CL.lImM] (clear Check limits).
6. Press No to use the stored check limits and proceed to step 9.
7. Press Yes to establish new check limits. The scale will then display [Set. Lo].
Press Yes to view the "Low" limit value. Press Yes to accept or No to edit
the "Low" limit value. The stored value then displays with the first digit
highlighted [000000 Pcs]. Repeatedly press No until the desired number
appears. Press Yes to accept and highlight the next digit.
Repeat until all the digits are correct. Press Yes to accept the "low" limit
value, [Set. Hi] will be displayed.
8. Repeat the same procedure to accept or edit the "high" value.
9.
If required, place an empty container on the pan and press Tare.
EN-9