Features and Functions
Measuring Capacitance
Disconnect circuit power and discharge all high-voltage capacitors before
CAUTION
measuring capacitance to prevent possible damage to the multimeter or the
device under test. To confirm that capacitors have discharged, use the DC
voltage function.
Measuring Temperature
– Do not bend the thermocouple leads at sharp angles. Repeated bending
CAUTION
– Do not contact the temperature sensor to any surface that is energized
36
over a period of time can break leads.
voltage or current sources, such the voltage source will pose a shock
hazard.
Measuring tips:
– For measuring capacitances greater than
10,000 μF, discharge the capacitor and
manually select a suitable measurement
range. This will speed up measuring time
in order to obtain the correct capacitance
value.
– Ensure the correct polarity when
measuring polarized capacitors.
– For measuring small capacitances,
press
subtract the residual capacitance of the
multimeter and leads.
Keysight U1241B and U1242B User's and Service Guide
with the test leads open to