Descargar Imprimir esta página

Keysight U1241B Guía Del Usuario Y Servicios página 64

Ocultar thumbs Ver también para U1241B:

Publicidad

Idiomas disponibles

Idiomas disponibles

5
Performance Tests and Calibration
Recommended Test Equipment
The test equipment recommended for the performance verification and
adjustment procedures is listed below. If the exact equipment is not available,
substitute the calibration standards of equivalent accuracy.
Table 5-1
Recommended test equipment
Application
Recommended Equipment
DC Voltage
Fluke 5520A
DC Current
Fluke 5520A
AC Voltage
Fluke 5520A
AC Current
Fluke 5520A
Resistance
Fluke 5520A
Frequency
Fluke 5520A
Capacitance
Fluke 5520A
Diode
Fluke 5520A
Fluke 5520A
Temperature
TM Electronics KMPC1MP
(K-Type thermocouple extension)
Short
Pomona MDP-S
64
Recommended Accuracy Requirements
<1/5 instrument 1 year specifications
<1/5 instrument 1 year specifications
<1/5 instrument 1 year specifications
<1/5 instrument 1 year specifications
<1/5 instrument 1 year specifications
<1/5 instrument 1 year specifications
<1/5 instrument 1 year specifications
<1/5 instrument 1 year specifications
<1/5 instrument 1 year specifications
<1/5 instrument 1 year specifications
Keysight U1241B and U1242B User's and Service Guide