Performance Tests and Calibration
Table 5-4
Calibration Adjustments (continued)
Step
Test Function
Press
to go to
9
function
Move the test lead from "μA.mA" and "COM" terminal to "A" and "COM" terminal
Caution: Connect the calibrator to the mul timeter's "A" and "COM" terminal before applying 10 A
Turn the rotary switch to
10
position
Press
to go to
11
function
Turn the rotary switch to
12
[a]
position
13
Press
to go to
[a] — Set the 5520A to internal reference.
— Prior to performing adjustment, connect one end of the K-type thermocouple (with miniature TC connector on both ends) to the 5520A TC
output, and the other end to a precision thermometer to verify that the source outputs the desired value. Adjust the source accordingly if
necessary.
— To perform the adjustment, connect one end of the K-type thermocouple (with miniature TC connector on both ends) to the 5520A TC
output, and the other end to the multimeter via a TC-to-banana adapter. Allow at least 1 hour for the multimeter to stabilize.
82
Cal Range
100 mA
mA
1000 mA
Open
10 A
A
10 A
Short
T1
T1T2
or
100 mV
T1
[a]
K-type
function
Keysight U1241B and U1242B User's and Service Guide
Input Reference Value
5 mA, 70 Hz
10 mA, 70 Hz
100 mA, 70 Hz
100 mA, 70 Hz
320 mA, 70 Hz
Input terminals open
(remove all test leads
and shorting plugs from
input terminals)
10 A
0.5 A, 70 Hz
1 A, 70 Hz
10 A, 70 Hz
Dual banana plug with
copper wires short
between two terminals
100 mV
0 ºC
Cal Item
U1241B
U1242B
5.00 mA
10.00 mA
100.00 mA
100.0 mA
320.0 mA
oPEn
10.000 A
0.500 A
1.000 A
10.000 A
SHrt
100.00 mV
000.0 ºC