Descargar Imprimir esta página

Keysight U1241B Guía Del Usuario Y Servicios página 47

Ocultar thumbs Ver también para U1241B:

Publicidad

Idiomas disponibles

Idiomas disponibles

Alerts and Warning During Measurement
Overload Alert
For your safety, please be aware of the alert. When you are alerted, just
WARNING
remove the test leads from the measuring source.
The multimeter provides an overload alert for voltage measurement in both auto
and manual range modes. The multimeter beeps periodically once the measuring
voltage exceeds 1100.0 V. For your safety, please be aware of this alert.
Input-A Warning Alert
The multimeter sounds an alerting beep when the test lead is inserted to the A
input terminal but the rotary switch is not set to the corresponding A location. The
display indicates a flashing AErr annunciator until the test lead is removed from
the A input terminal. This warning alert is not available in T1/T2 temperature
measurements mode.
Input-mA Warning Alert
The multimeter sounds an alerting beep when the μA/mA input terminal detects a
voltage level of more than 1.6 V. The display indicates a flashing CErr annunciator
until the test lead is removed from the μA/mA input terminal.
Keysight U1241B and U1242B User's and Service Guide
Features and Functions
2
47

Publicidad

loading

Este manual también es adecuado para:

U1242b