Controller
Before the appliance is loaded with vac-
cines ensure that the temperature is below
-15 °C:
● Set the thermostat at position 5
● Check the temperature on the thermo-
meter
● If the temperature in the vaccine com-
partment is above -15°C the thermostat
should be adju sted. The temperature in
the vaccine compartment must always
be monitored on the thermo meter and be
within the range -15 to -25°C.
Due to the mechanical tolerance of the
thermostats you should always control the
temperatu re during cool down .
Temperature adjustment
This appliance is equipped with one
thermostat.
Thermostat
The thermostat is a electronic thermostat
that controls the temperature in the vaccine
and icepack freezing compartments.
Setting 1 applies minimum cold and setting
8 maximum cold.