On-site checklist
● Is the control panel on (Power check)
● Temperature records ( manual records,
FT2 data)
● Is the internal temperature inside the
acceptable range of -15° to -20°
● Is the vaccine compartment clean and
without ICE BUILD UP more then 0,5 cm
on inner lining
● Is the Compressor running
● Is the appliance placed according to
instruction in the manual.
● Does the lid close tight to cabinet and is
the lid gasket in good condition
● Is the grille for compressor compartment
clean
● Is all electrical components working
properly
● Over all condition of the cabinet –internal
and external: any corrosion, rusting,
cracks
● Inspection of the refrigeration line
(the condenser, evaporator, the whole
refrigeration circuit/line)