Pulse oximeter
FR
IT
DE
PT NL
PL
DK
ES
EN
SWE
A. Alarm setting
In the main menu interface, move the choice bar to "Alarm" item, then press "right button" to enter the alarm setting
menu of Figure 7:
Figure 7. Alarm setting menu
a. The high/low limit of alarm setting:
In alarm setting menu,you can set the high/low limit of alarm.Move the choice bar to the menu item that you want to
set, and press "right button" to begin to set value.Press "menu button" for affirming.
If the alarm function is on, the device will provide medium-priority alarm signal when the data of SpO2 or pulse
rate is beyond the limit. Intermittent alarm will occur and the value shows in yellow. Medium priority indicating that
prompt operator response is required.
b. The alarm setting
In alarm setting menu,move the choice bar to the "Alarm" item, then set the alarm state by pressing "right button".
Choose "on" to turn on the alarm, and choose "off" to turn off the alarm.
c. Pulse sound indication setting
In alarm setting menu,move the choice bar to the "Pulse Sound" item, then press "right button" to set state, choose "on"
to turn on pulse sound, and choose "off" to turn off pulse sound.
B. Clock setting
In the main menu interface, move the choice bar to "Clock" item, then press the "right button" to enter the clock setting
menu of Figure 8:
Figure 8. Clock setting menu
27