Under trend review mode as Figure 15,the white number on the left bottom is current recording time point, the middle
blue number is SpO2 value,green number is pulse rate value,the white number on the right bottom is PI value.Press "up
button" or "down button" to page up or page down;press "left button" or "right button" to move the red triangle.
H. Close the device
a In the main menu interface, move the choice bar to the "Power off' item,then press menu button to close the device.
If the storage function has been opened,the prompt interface of "Recording... " will appear when closing the device.It
means that the device is in the record state,can't be closed.
b In the state of boot-strap,long press power button could close the device too.
If "finger out" appears and you don't turn on the record function and there's no operation in 10 minutes, the device
will close automatically.
6.1.4. PC software operation
Please connect the device to the computer by USB data line ,which is affiliated with the device, then double click "SpO2
Assistant" icon to run the PC software. The functions such as uploading record data and change device ID could be
carried out by the software.Please refer to <SpO2 Assistant user manual> for detail.
If the users choose to turn on the synchronizing display function on computer, it would probably take several
seconds for the data to appear on the computer screen.(If there is no data on the computer screen ,unplug USB data
line,then repeat step t 6.1.4"again).
6.1.5. Charge
There are two kinds of charge method:
A. Connect the device to computer with data line, then the device should be in charge state.
B. Connect the device to power supply with power adaptor, then the device should be in charge state.
When pressing power button,if the device is closed,the charge icon will display,and when the battery status is full,the
icon of full power will display.If there is no operation in 30 seconds,the display will closed automatically, here short
press power button to turn on display. When charging,long press power button to turn on the device,here the battery
status icon on the right top will display dynamically.It means that the device is charging up. When the battery status is
full, the icon of full power will display.
If the alarm function is on, the device will provide high-priority alarm signal when the battery is in low power
status. Intermittent alarm will occur. High priority indicating requires that operator responds immediately.
DE
PT NL
SWE
Figure 15. Trend review mode
PL
DK
Pulse oximeter