• The system must not be used adjacent or supported by other
equipment. The recommendations of this manual must be
followed.
• Do not use accessories, transducers, internal parts of
components and other cables other than those previously
specified by the manufacturer. This may result in increased
emission or decreased electromagnetic immunity and result in
improper operation.
• Portable RF communications equipment (including peripherals
such as antenna cables and external antennas) should by used
no closer than 30cm to any part of the ultrasound system,
including cables specified by the
manufacturer. Otherwise, degradation of the performance of this
equipment could result.
• To maintain basic safety in relation to electromagnetic
disturbances during the expected service life, always use the
system in the specified electromagnetic environment and follow
the maintenance recommendation described in this manual.
The following tables provide information on compliance
of the equipment according to the standard
EN 60601-1- 2:2015.
Table 1. Compliance class
Emissions
Test
RF emissions
CISPR 11
RF emissions
CISPR 11
Compliance
Group 1
Class B
Electromagnetic Environment and
Guidance
The equipment uses RF energy only
for its internal function. Therefore, its
RF emissions are very low and are
not likely to cause any interference
in nearby electronic equipment.